<What is PPG?>
- PPG
> Optical technique used to detect volumetric changes in blood in peripheral circulation
→ Low cost & non-invasive method (Measurements at the surface of the skin)
→ Measurable list: Oxygen saturation, Blood pressure, Cardiac Output, etc.
- Principle
> When heart beats, the blood vessels are slightly inflated and arterial blood flow changes by pressure. Accordingly, light transmittance is measured → PPG signal
> Shine the LED light on human tissue which includes microvascular and detect the difference of light by using photo detector
<Charateristic of PPG>
- (Quasi) DC + AC
> PPG waveform comprises a pulsatile physiological waveform attributed to cardiac synchronous changes in the blood volume with each heart beat, and is superimposed on a slowly varing baseline with lower frequency components attributed to respiration, sympathetic nervous system activity and thermoregulation
> Average respiration rate of adult: 20 per minute → About 0.3~0.4 [Hz] (Breathing Rate)
> Average heart rato of adult: 60 ~ 120 per minute → About 1~2 [Hz] (Heart Rate)
- PPG signal characteristic by light wavelength
> Green: The light wavelength where hemoglobin absorb most compared to other lights (Great absorptivity) -> Clearly distinguished PPG signal is detected and commonly used for measuring heart rate
> IR (Infrared Ray): Less effected by (skin) color -> used in wear detection. But, very vulnerable to motion effect. Also, by measuring red and IR simultaneously and calculating light transmittance difference of oxy, de-oxy hemoglobin, we can detect oxygen saturation.
<Circuit of PPG>
- Main device: LED (Light Emitting Diode), PD (Photo Detector), TIA (Transimpedance Amplifier)
> TIA: For amplifying PD current because PD current is small
<Drawback of PPG>
- Vulnerable to Motion: MA (Motion Artifact)
1. Motion result in the leakage of lights → Directly affect to PPG
2. Deform the microvascular system
→ But, the frequency range of PPG (0.5~4.0 [Hz]) is considerably overlap to the frequency range of MA (0.01~10 [Hz]) ∴ Normal filtering method for eradicating MA noise couldn't be applied
→ Many algorithm research for removing MA noise are in underway
'논문 및 소자 리뷰' 카테고리의 다른 글
[E-Drug] Further study: Gas permeable membrane (0) | 2022.01.11 |
---|---|
E-Drugs (Electronic Drugs): Spatial and Temporal Medical Treatment of Human Disease (0) | 2022.01.11 |
Voltage Regulator (전원 레귤레이터) (0) | 2021.07.21 |
ECG (Electrocardiography) Measuring Circuit [Programming] (0) | 2021.07.10 |
ECG (Electrocardiography) Measuring Circuit (0) | 2021.07.10 |